RecallDepth

Plastic Cannula Holdex Tube Holder, Item number: 450216. It is intended for use in routine specimen withdrawal from specimen collection bags or bottles through needleless cannula ports.

Company
Greiner Bio-One North America, Inc.
Recall Initiated
August 5, 2008
Posted
September 30, 2008
Terminated
July 25, 2012
Recall Number
Z-2198-2008
Quantity
150,000 pcs
Firm Location
Monroe, NC

Reason for Recall

At removal of the tube from the holder the needle may dislodge and become stuck in the tube stopper with the blunt end of the needle facing out.

Distribution

AL, FL, IN, KS, KY, MO, MN, NC, OH, PA, TN, TX, UT, VA, WI and WV.

Lot / Code Info

Lot#: A030804

Root Cause

Process design

Action Taken

Consignees were notified by letter on 08/06/2008. Letters to Distributors instructed them to stop distribution of the affected product immediately and to hold any product of the lots they may have pending returned to Greiner bio-one. The were instructed to provide Greiner bio-one with a list of their customers who may have received the affected lot so that they can contact them. A Product Disposition form was included to be completed and returned to Greiner bio-one. Contact Greiner Bio-One North America, Inc. at 1-888-286-3883 for assistance.