Smooth or threaded metallic bone fixation fastener. 3.5mm x 14.0 mm Locking Cortical Screw. Product labeled in 3.5mm x 14.0 mm Locking Cortical Screw. Product labeled in part, "Congruent Plate System... Size: 3.5mm x 14.0 mm Locking Cortical Screw... REF COL-3140-S LOT W46772... Acumed 5885 NW Cornelius Pass Road Hilsboro, OR ". Product is used as an orthopedic implant.

This recall has been terminated (originally issued September 11, 2008).

Company
Acumed Llc
Recall Initiated
July 9, 2007
Posted
September 11, 2008
Terminated
February 3, 2010
Recall Number
Z-2135-2008
Quantity
400 screws
Firm Location
Hillsboro, OR
Official Source
View on FDA website ↗

Reason for Recall

Product mislabeled. The inner sticky labels for the patient identified the product incorrectly as a Cortical screw (non-locking). The product package was labeled correctly.

Distribution

Class III Recall - Worldwide Distribution --- USA including countries of UK and Japan.

Lot / Code Info

Part number: COL-3140-S; Work order number/Lot: W46772 (W=work order, ###=sequential number); Expiration date: 2012-02 (Year/Month)

Root Cause

Packaging process control

Action Taken

The firm issued a Product Removal notification on 8-9-2007. The notification was reportedly sent by either FAX or e-mail. The firm notified consignees of the issue and gave the consignees the choice of returning the screws. The notification did not include any instructions for a sub-recall. If you have any questions contact Jason Gozdecki at 1-888-627-9957.