RecallDepth

iReview" Software (Accessory To) iLab Ultrasound Imaging System, Model Number: H74900018010, Catalog Number: 1801; Product is manufactured and distributed by Boston Scientific Corporation, Fremont, CA The iReview" Software is provided on CD and is a computer application designed for reviewing cases recorded with firm's iLab "System, and for exporting stills or videos of those runs.

Company
Boston Scientific Corporation
Recall Initiated
May 2, 2008
Posted
September 17, 2008
Terminated
June 1, 2009
Recall Number
Z-2006-2008
Quantity
10 kits
Firm Location
Fremont, CA

Reason for Recall

The firm initiated a recall of a specific iReview" Software, Ver 1.0 Kit, which was released for distribution in a non-validated format.

Distribution

Product was distributed in FL, MI, CA, AZ and SC

Lot / Code Info

Lot Number: 74791

Root Cause

Nonconforming Material/Component

Action Taken

Customer notifications sent on May 2, 2008 indicating the need for field correction and included a replacement version of the software kit. For customer assistance, contact Boston Scientific at 1-508-652-5594.

More recalls by Boston Scientific Corporation