Boston Scientific mach 1 Guide Catheter, VL3.5 SH Voda Left 3.5, REF 34358-78, UPN H74934358780, Sterile EO, Made in Mexico Avenida Norske Edicficio G1, Local B y H Parque Industrial La Mesa, Fracc. Rubio La Mesa, Tijuana, B.C. Mexico C.P. 22550

This recall has been terminated (originally issued October 27, 2007).

Company
Boston Scientific
Recall Initiated
September 11, 2007
Posted
October 27, 2007
Terminated
February 24, 2008
Recall Number
Z-0104-2008
Quantity
7
Firm Location
Maple Grove, MN
Official Source
View on FDA website ↗

Reason for Recall

Missing holes: Catheters are missing the catheter's side holes.

Distribution

Worldwide: NY, PA, and Japan

Lot / Code Info

Lot/Batch # 559022

Root Cause

Process control

Action Taken

A voice mail was left to affected members of the domestic sales force on September 11, 2007 and an interoffice memo with the recall implementation instructions was also sent to the Sales Force. An Urgent Voluntary Medical Device Recall letter dated September 11, 2007, was sent to affected consignees describing the problem and products affected. The letter included the customer steps for recall, a Reply Verification Tracking Form and a return shipping label.