Remington Medical, Inc, (RMI) -- Urinary Drainage Bags: 500ml Disposable Urinary Leg Bag (Order number: 500-D) and 1000ml Disposible Urinary Leg Bag (Order number 1000-D) -- Sterile Fluid Path, For Single Use Only, Remington Medical, Inc., Alpharetta, GA 30005.

This recall has been terminated (originally issued October 23, 2007).

Company
Remington Medical Inc.
Recall Initiated
July 23, 2007
Posted
October 23, 2007
Terminated
April 6, 2009
Recall Number
Z-0058-2008
Quantity
45,000 units
Firm Location
Alpharetta, GA
Official Source
View on FDA website ↗

Reason for Recall

Sterility, inadequate -- Pathway may not be sterile as labeled.

Distribution

Nationwide Distribution -- including states of AL, AR, AZ, CA, CO, CT, GA, HI, IL, IN, KS, KY, MD, MI, MT, NB, NC, OH, NY, PA, RI, TN, WA, WV, and WY.

Lot / Code Info

Lot # 20070423 (500-D), and Lot #20070420 (1000-D)

Root Cause

Process control

Action Taken

Consignees were notified by Urgent Recall Letter (dated 07/23/07) via fax from 07/24/2007 through 07/27/2007. They were instructed to quarantine all affected product in inventory. If product was distributed they were instructed to identify their customers and notify them of the recall via copy of the notification letter.