3D TOP Ceiling Stand. X-Ray System model number 8773673
This recall has been terminated (originally issued December 31, 2005).
- Recall Initiated
- November 30, 2005
- Posted
- December 31, 2005
- Terminated
- July 19, 2006
- Recall Number
- Z-0335-06
- Quantity
- 460 units
- Firm Location
- Malvern, PA
- Official Source
- View on FDA website ↗
Reason for Recall
screws on the rollers may become loose
Distribution
The products were shipped to hospitals and medical facilities nationwide.
Lot / Code Info
Model number 8773673
Root Cause
Other
Action Taken
The recalling firm has issued a field correction per Update Instructions AX004/06/S. This update instructs a Siemens Customer Service Engineer to visit affected consignees and checkthe screws on the rollers. It also provides roller bearing sleeves as a secondary safety mechanism to prevent the rollers from detaching.