RecallDepth

Stainless Steel Greenfield Vena Cava Filter with 12 Fr / 4,0 mm FlexCarrier Capsule. For Femoral Vein Introduction Only. Order No. 50-501. Sterile EO. Single Use Only. Read Instructions for Use before using this device. Distributed in USA by: Boston Scientific Corporation, One Boston Scientific Place, Natick, MA USA 01760. Manufacturer: Boston Scientific Cork Ltd., Business & Technology Park, Model Farm Road, Cork, Ireland. The Stainless Steel Greenfield Vena Cava Filter with 12 Fr. / 4,0 mm introducer System is a permanently implanted stainless steel device designed to protect against pulmonary embolism while maintaining patency of the inferior vena cava. The Stainless Steel Greenfield Vena Cava Filter comes preloaded in a 12 Fr. / 4,0 mm jugular or femoral introducer catheter.

This recall has been terminated (originally issued December 20, 2005).

Company
Boston Scientific
Recall Initiated
December 2, 2005
Posted
December 20, 2005
Terminated
April 2, 2007
Recall Number
Z-0280-06
Quantity
18,000
Firm Location
Maple Grove, MN
Official Source
View on FDA website ↗

Reason for Recall

There have been reports of detachment at the bond between the carrier capsule and the outer sheath of the Greenfield Vena Cava Filters with 12 Fr Femoral introducer Systems manufactured before Marcyh 10, 2004. If the capsule should detach during an implantation procedure, there is a risk of cardiac and pulmonary embolization.

Distribution

nationwide (except ND); Virgin Islands, OUS to include: united Arab Emirates, Argentina, Australia, Belgium, Brunei Darussalam, Brazil, Canada, Chile, China, Colombia, Germany, Spain, France, Ireland, Great Britain, Israel, India, Italy, Jamaica, Jordan, Japan, Korea (South Republic), Kuwait, Lebanon, Mexico, Malaysia, Netherlands, New Zealand, Panama, Philippines, Poland, Portugal, Saudi Arabia, Singapore, South Africa, Thailand, Trinidad and Tobago, Taiwan, Venezuela

Lot / Code Info

all codes of product manufactured before March 10, 2004. Catalog no. 50-501 with lot/batch # between 5145758 and 6387904

Root Cause

Other

Action Taken

Recall packages started going out 12/02/05 via Fed Ex and was completed 12/03/05. A letter was directed to several persons within each hospital . The letter describes the issues, the affected lot numbers and the complaints that Boston Scientific has received. Affected devices are to be segregated for return back to Boston Scientific. A 'Reply Verification Tracking Form' is asked to be returned back to Boston Scientific.

More recalls by Boston Scientific

View all recalls by this company →