0.10% BreathScan Alcohol Detector manufactured for WNCK, Inc., The Woodlands, TX by Howard Industries, Inc., Columbus, OH (crystals) and James Alexander Corp., Blairstown, NJ (general device assembly). Product sold in bulk and individually packaged. Bulk product (500-1000 units) distributed in unlabeled clear plastic bags. Individually packaged product packaged and sealed in clear plastic bags with labeling and sold in boxes containing 100 units/box.

This recall has been terminated (originally issued December 23, 2005).

Company
Wnck, Inc
Recall Initiated
November 30, 2005
Posted
December 23, 2005
Terminated
April 25, 2007
Recall Number
Z-0323-06
Quantity
126,552 total units.
Firm Location
The Woodlands, TX
Official Source
View on FDA website ↗

Reason for Recall

Device marketed without 510(k) as required; 0.10% BreathScan Alcohol Detector turning positive with breath alcohol samples 50% below the test cutoff.

Distribution

Nationwide; product distributed to distributors, retailers and consumers/users in CA, CO, FL, LA, MN, MS, NY, OH, PA, TN, TX, VA, WA. Product also distributed to Markham, Ontario, Canada and Kfar Saba, Israel.

Lot / Code Info

D121898

Root Cause

Other

Action Taken

The firm initiated recall on 11/29/05 via letter to all distributor consignees.