The BABIX holder is an accessory used with the BuckyDiagnost X-ray system. It is used to hold the baby either in a secured or immobilized standing position in front of the X-ray detector during the examination.

This recall has been terminated (originally issued October 7, 2010).

Company
Philips Medical Systems North America Co. Phillips
Recall Initiated
September 9, 2010
Posted
October 7, 2010
Terminated
July 22, 2011
Recall Number
Z-0021-2011
Quantity
4 units
Firm Location
Bothell, WA
Official Source
View on FDA website ↗

Reason for Recall

Potential of broken strands in the main steel cable supporting the accessory holder used with BuckyDiagnost and DigitalDiagnost X-ray systems.

Distribution

Nationwide Distribution in the states of CA, GA, and TN.

Lot / Code Info

Devices are identified as Site Numbers: 546607, 48682949, 554660, and 555738.

Root Cause

Equipment maintenance

Action Taken

Philips issued URGENT-Device Correction Notice letters dated September 9, 2010 to their consignees identifying the affected products and action to be taken by customers. The letter states that Babix holder used with Philips BuckyDiagnost and DigitalDiagnost X-Ray System - VS Wall Stand Risk of steel cable fracture in the support holder. The consignees are informed that there is a potential of broken strands at the main steel cable within the holder which cannot be recognized by the user. The firm will issue the Field Change Order (FCO) and the Service Engineer will replace the Babix holder. Consignees can call Philips Call Center at 800-722-9377, #5 (then follow prompts) and reference to FCO 70400038 for the BuckyDiagnost X-ray system and FCO 71200046 for the DigitalDiagnost X-ray system.