Enseal TRIO Product Code ETRIO335H, Lot Code: G4TC14, Exp. Date 2012/05

This recall has been terminated (originally issued September 30, 2010).

Company
Ethicon Endo-Surgery Inc
Recall Initiated
July 9, 2010
Posted
September 30, 2010
Terminated
February 14, 2011
Recall Number
Z-2635-2010
Quantity
396 instruments
Firm Location
Cincinnati, OH
Official Source
View on FDA website ↗

Reason for Recall

The weld that connects the electrode to the active rod was separated. This connection is critical to the proper function of these devices. If the weld is separated, the product will not deliver energy to the jaws, resulting in the potential for a cut/transection without a seal. Ethicon Endo-Surgery identified this issue during a routine finished goods quality inspection.

Distribution

Nationwide Distribution to the following states: AZ, AL, AZ, CA, CO, CT, FL, GA, IN, KY, MA, MD, MI, MO, MS, NC, NV, NY, OH, PA, TN, TX, WA, and WV.

Lot / Code Info

Product Code ETRIO335H, Lot Code: G4TC14, Exp. Date 2012/05

Root Cause

Other

Action Taken

Ethicon Endo-Surgery issued an Urgent: Device Recall, Event 2193 letter dated July 19, 2010 identifying the affected device, the issue prompting the recall, and actions to be taken by consignees. Consignees were instructed: Do not use the affected product. Identify and quarantine affected product from inventory. Complete the Business Reply form and fax to Stericycle at 877-597-9580 within three days To return product, place it in a box with a copy of the Business Reply form and mail to Stericycle using the pre-printed UPS shipping label. Replacement product should be received by the customer within 60 days of return of affected product. Customers can contact Stericycle about this recall at 877-243-7314 and referencing Event 2193.