Medtronic Everest 20 Disposable Inflation Device, AC2200. Used to inflate/deflate balloon catheters as well as to monitor pressure within the balloon.

This recall has been terminated (originally issued August 25, 2010).

Company
Medtronic, Inc.
Recall Initiated
July 16, 2010
Posted
August 25, 2010
Terminated
June 7, 2011
Recall Number
Z-2271-2010
Quantity
2,875 units
Firm Location
Danvers, MA
Official Source
View on FDA website ↗

Reason for Recall

Product sterility may be compromised due to breach of package seal integrity.

Distribution

Worldwide Distribution -- USA, including the state of MI and the countries of Austria, Belgium, Czech Republic, Finland, France, Germany, Greece, Ireland, Israel, Italy, Lithuania, Pakistan, Poland, Portugal, Slovakia, Spain, Sweden, Switzerland, Ukraine, United Arab Emirates, and United Kingdom.

Lot / Code Info

Lot numbers: 50517330, 50517331,50517332, 50517333, 50517334, 50519716, 50519717, 50519718, and 50519679.

Root Cause

Packaging process control

Action Taken

Medtronic issued a Urgent Medical Device Recall Notice to US customer on 7/15/2010 and to foreign accounts advising of the issue and requesting that they segregate recalled product. A Medtronic Sales Rep will visit their facility to account for all affected product and initiate a credit. Foreign accounts will return product to their respective distribution centers.