RecallDepth

Hallmark Anterior Cervical Plate System: Primary Bone Screw, Semi Constrained, REF # 62-3140,14mm; Blackstone Medical, 90 Brookdale Drive, Springfield, MA 01104 (The Hallmark Anterior Cervical Plate Systems consists of an assortment of implantable titanium alloy plates and bone screws that are provided non-sterile)

This recall has been terminated (originally issued January 26, 2008).

Company
Blackstone Medical, Inc.
Recall Initiated
May 8, 2006
Posted
January 26, 2008
Terminated
November 6, 2008
Recall Number
Z-0478-2008
Quantity
1,991 semi-constrained primary and rescue screws.
Firm Location
Wayne, NJ
Official Source
View on FDA website ↗

Reason for Recall

Incomplete Design Validation: the laser marking on the head of the screw lacked complete process qualification

Distribution

Nationwide

Lot / Code Info

All Lots (Primary semi-contained rescue Screws: P/N's 62-3100 through 62-3180 and rescue simi-constrained bone screws P/N's 62-4100- 62-4180)

Root Cause

Device Design

Action Taken

On May 8, 2006, a communication to the field was sent via phone, e-mail and fax. Instruction was made to retrieve and return all product in the field.