RAD SOURCE RS 3000 Blood Irradiator Three Bag Option

This recall has been terminated (originally issued July 20, 2004).

Company
Rad-Source Technologies, Inc. 20283 State Rd. 7 Suite 107 Boca Raton Fl 33498
Recall Initiated
June 1, 2002
Posted
July 20, 2004
Terminated
May 3, 2012
Recall Number
Z-0745-04
Quantity
20
Official Source
View on FDA website ↗

Reason for Recall

The firm completed a cooling system retrofit to preclude overheating and failure.

Distribution

Units were distributed to 20 blood banks and transfusion services in WI, KS, IN, MI, GA, NC, WA, NE, FL, IN, TX, MD, Washington DC, SC, and NM.

Lot / Code Info

Serial numbers: 1002 through 1005, 1007 through 1009, 1011 through 1023.

Root Cause

Other

Action Taken

The firm e-mailed a notification on May 9, 2002 to its consignees giving them instructions on using additional radiation indictors to assure that the blood units were correctly irradiated. They also initiated a cooling tube retrofit of all units on that date and this was completed on 8/13/2002.