RecallDepth

Exel Safety Huber Infusion Sets (20G and 22G sizes with Catalog Numbers 37854S, 37855S. 37858S).

This recall has been terminated (originally issued January 22, 2010).

Company
Exelint International Co.
Recall Initiated
December 23, 2009
Posted
January 22, 2010
Terminated
June 6, 2011
Recall Number
Z-0639-2010
Quantity
3,360
Firm Location
Culver City, CA
Official Source
View on FDA website ↗

Reason for Recall

Coring of port's septum, which may lead to: Decreased effectiveness of port, infection, necrosis, swelling, and embolization, which may result in Pulmonary Embolism, Limb ischemia, stroke, local extravasation of medication such as chemotherapy or other caustic material, may lead to local tissue necrosis, granulomas, failure to deliver therapy, need for additional surgery to replace implanted port

Distribution

US (VA TX NV KY OH NH MS NY OH NM DE WA AZ IL PA OR CO FL GA LA UT MI WI NV CA NJ MO IA) and Canada.

Lot / Code Info

Catalog 37854S Lot 09G31; Catalog 37855S Lot 09G31, Catalog 37858S Lot 09G31.

Root Cause

Device Design

Action Taken

Distributors have been advised to cease distribution of any recalled needles and infusion sets; to quarantine any remaining; to document amount on Customer Response Form; to return form to Exel (firm); to arrange for return of product to Exel for credit. Distributors also are advised to share recall notice with all affected customers. Contact phone: 727-827-1922x0 (M-F 9a-5p EST). Contact email: info@exelint.com

More recalls by Exelint International Co.

View all recalls by this company →