RecallDepth

Accolade System Surgical Technique; Accolade C - Cemented Hip Stem; Accolade TMZF Cementless Hip System; Accolade HFx Cementless Hip System. Stryker Orthopaedics. Surgical protocols instruct surgeons on proper surgical technique.

This recall has been terminated (originally issued January 11, 2010).

Company
Stryker Howmedica Osteonics Corp.
Recall Initiated
August 31, 2009
Posted
January 11, 2010
Terminated
July 17, 2013
Recall Number
Z-0504-2010
Quantity
5,670
Firm Location
Mahwah, NJ
Official Source
View on FDA website ↗

Reason for Recall

Certain surgical protocols do not provide specific instructions for initiating power to the calcar planer prior to engagement with bone.

Distribution

Worldwide Distribution.

Lot / Code Info

Catalog Number/Lot Code: LASST.

Root Cause

Error in labeling

Action Taken

"Urgent Product Correction - Surgical Protocol" letters and Product Accountability Forms were sent to consignees via Fed Ex on August 31, 2009 and September 1, 2009 with return receipt. Notification letters described the issue, protocols affected, potential hazard and harm and risk mitigation. Direct questions about the recall by calling 1-201-972-2100 or Customer Service 1 -866-OR-ASSIST (1 -866-672-7747).

More recalls by Stryker Howmedica Osteonics Corp.

View all recalls by this company →