SJM Peel Away Introducer, REF 405153, Lot 2694252, Length 14 cm, 7F (2.33 mm), Max Guidewire o.d .038" (.965 mm), contents 5, 2011-08, Manufacturer St Jude Medical 14901 DeVeau Place, Minnetonka, MN 55345-2126 USA, Sterile EO, Rx only. Intended for use for the introduction of catheters into a vessel.

This recall has been terminated (originally issued August 17, 2009).

Company
St. Jude Medical
Recall Initiated
June 24, 2009
Posted
August 17, 2009
Terminated
March 29, 2012
Recall Number
Z-1714-2009
Quantity
200
Firm Location
Minnetonka, MN
Official Source
View on FDA website ↗

Reason for Recall

St. Jude Medical has initiated a voluntary recall of two lots of Peel Away Introducers . Isolated lots could contain some items that have been mis-labelled with the incorrect French size of the outer box label. Some 9F products have been labelled as 7F and vice versa. Only the outer box label is affected by this error, the labels on the individual product pouches correctly identify the product. St Jude Medical has not received any reports of injures resulting from this mislabelling to date.

Distribution

OUS only: Belgium.

Lot / Code Info

Lot Number: 2694252.

Root Cause

Labeling mix-ups

Action Taken

The Consignee was provided with a St. Jude Medical "Urgent Product Information" letter and a doctor Acknowledgement form, via e-mail on June 2009. The letter described the problem and the product involved and advised to isolate and discontinue use of the product involved. The Consignee will notify the end user and remove any remaining affected units and will notify St Jude Medical of the results.