9051 Electrode Adapter; Automated external defibrillator, Used with QUIK-COMBO System; Cardiac Science Corp, Bothell, WA 98021.

This recall has been terminated (originally issued May 21, 2008).

Company
Cardiac Science Corporation
Recall Initiated
January 23, 2008
Posted
May 21, 2008
Terminated
July 21, 2011
Recall Number
Z-1616-2008
Quantity
Lot W029275 - 24; Lot W029475 - 38
Firm Location
Bothell, WA
Official Source
View on FDA website ↗

Reason for Recall

Adaptor Cable Incorrectly Manufactured: The red and white connector ends of the adapter are switched. Therefore, the cable will not connect to the electrodes as indicated in the instructions.

Distribution

Nationwide Distribution including states of CT, IA, IN, NC, NY, OH, PA, WA, and WI.

Lot / Code Info

Lot Numbers: W029275 and W029476.

Root Cause

Process control

Action Taken

Consignees were notified by an Urgent Medical Device Recall letter sent on 01/23/2008. The letter instructs users to check each cable of the affected lot to ensure that it properly connects to the electrode. Users were also instructed to return any defective product to the recalling firm for a replacement. For additional information, contact 1-888-466-8686 x2041.