RecallDepth

Wasatch Photonics 785/830-SR and 785-ER-ILC Raman Spectrometers, Product Models: 785-SR-ILP, 785-SR-ILC, 830-SR-ILP, 830-SR-ILC,785-ER-ILC.

This recall is currently active, issued September 11, 2025. It was issued by Wasatch Photonics.

Company
Wasatch Photonics
Recall Initiated
August 4, 2025
Posted
September 11, 2025
Recall Number
Z-2478-2025
Quantity
192
Firm Location
Morrisville, NC
Official Source
View on FDA website ↗

Reason for Recall

Two new potential laser safety failure modes have been identified in the product design. Overcurrent of laser diode, and a fail-unsafe mode. The failure modes have never been found to occur in these models, whether in the field or factory, but could potentially occur under excessive force or extreme, continuous vibration.

Distribution

U.S.

Lot / Code Info

all codes included in the model numbers.

Root Cause

Radiation Control for Health and Safety Act

Action Taken

Notifications will be sent out to all purchasers of affected models. Notification will include an offer of free RMA (return to factory for repair, then shipment back to customer). All repairs will be performed at the site of original manufacture in Morrisville, NC. Customer notifications to be delivered by 31-Aug-2025.