OPUS SpeedStitch Suture Cartridge with Magnumwire, a component of the OPUS MiniPlus Implant Set, for use in placement of sutures through soft tissue in endoscopic and other limited access procedures.

Company
Arthrocare Corporation
Recall Initiated
June 28, 2006
Posted
August 2, 2006
Terminated
December 5, 2007
Recall Number
Z-1312-06
Quantity
198 units
Firm Location
Sunnyvale, CA

Reason for Recall

The product may lose sterility due to omission of the final pouch seal.

Distribution

Nationwide.

Lot / Code Info

All products consisting of lot number 110399, Catalog number OM-6006. This catalog number is referred to as the OPUS MiniPlus Implant Set.

Root Cause

Packaging process control

Action Taken

The firm has issued notification letters to all customers in the United States on June 29, 2006 via overnight delivery, advising customers to return all product with incomplete or missing seals to Arthocare.