RecallDepth

Brand Name: AART Malar Implant Product Name: Malar Implant Model/Catalog Number: 401-202-5 Malar Implant style 2 size 2 left UDI-DI code: B4904012025 401-202-6 Malar Implant style 2 size 2 right UDI-DI code: B4904012026

Company
Dsaart Llc
Recall Initiated
September 26, 2025
Posted
December 18, 2025
Recall Number
Z-0947-2026
Quantity
2 implants
Firm Location
Carson City, NV

Reason for Recall

Due to manufactured products that deviated from established specifications and being distributed without obtaining the required FDA premarket approvals or clearances.

Distribution

U.S.: AR, AZ, CA, CO, CT, DC, FL, GA, KY, IL, MA, MD, MI, MN, MS, NE, NJ, NY, OH, OK, PA, PR, TX, UT, VA, WA, and WI O.U.S.: N/A

Lot / Code Info

Malar Implant Model/Catalog Number: 401-202-5 UDI-DI code: B4904012025 Lot Number: 24-01-014 Model/Catalog Number: 401-202-6 UDI-DI code: B4904012026 Lot Number: 24-01-015

Root Cause

No Marketing Application

Action Taken

On 09/26/2025, the firm emailed a notice to an affected customer informing them that due to identified testing issues, the firm was requesting return of gluteal implants. On 11/17/2025, the firm send an "URGENT Medical Device removal of AART Silicone Implants" Letter to customer informing them that Alpha Aesthetics is voluntarily removing products as they conduct an analysis to determine the regulatory status of the devices. Customers are instructed to: 1. Either return the affected products that have not been implanted to Alpha Aesthetics at the company's expense, or discard affected products. 2. Acknowledge their actions via the enclosed "Medical Device Return Response" letter/form. For questions or further assistance with this removal, contact ** or email **.