Frontier Devices, REF: 301.914S1, 14 mm Distraction Pin, 10 single packs , Rx Only, Sterile

This recall is currently active, issued September 17, 2025. It was issued by Folsom Metal Products, Inc..

Company
Folsom Metal Products, Inc.
Recall Initiated
August 21, 2025
Posted
September 17, 2025
Recall Number
Z-2579-2025
Quantity
10,960 units
Firm Location
Pelham, AL
Official Source
View on FDA website ↗

Reason for Recall

Labeling includes shelf life that has not been validated.

Distribution

US Nationwide distribution in the states of AL, CA, FL, GA, IL, MD, MI, PA, TX.

Lot / Code Info

All lots distributed from August 1, 2020, to July 30, 2025/UDI: 00850014548032

Root Cause

Incorrect or no expiration date

Action Taken

On August 21, 2025, Frortier Devices issued a "Urgent: Medical Device Recall" Notification to affected consignees via FedEx. On August 22, 2025, Frontier reissued the notification to correct an incorrect product number. Frontier asked consignees to take the following actions: 1. Immediately examine your inventory and quarantine any affected units. 2.If you have affected units: Return product to Frontier Devices at the address listed, OR Destroy product on-site and complete the enclosed Certificate of Destruction Form. 3. If you may have further distributed this product, please identify your customers and notify them at once of this product recall. Your notification to your customers may be enhanced by including a copy of this recall notification letter. Please share this notice with all users within your organization. If the product has been transferred to a different facility, please make them aware of the notice. 4. Complete and return the enclosed Medical Device Recall Response Acknowledgment Form as soon as possible, even if no affected product remains in your possession. Please have them complete the acknowledgement response form and return it to you.

More recalls by Folsom Metal Products, Inc.

View all recalls by this company →