Automated Impella Controller (AIC) with the below brands and product codes. Used with heart assist pump. 1. Impella Controller, Packaged, US; Product Code: 0042-0000-US. 2. Impella Controller, Packaged, US, Loaner; Product Code: 0042-0000-US-L. 3. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 4. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 5. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 6. Impella Controller, UK, Loaner; Product Code: 0042-0000-UK-L.

Company
Abiomed, Inc.
Recall Initiated
August 20, 2025
Posted
September 19, 2025
Recall Number
Z-2568-2025
Quantity
71 units
Firm Location
Danvers, MA

Reason for Recall

Specific Automated Impella Controllers (AICs) that have a Pump Driver Circuit Assembly do not meet current specifications. This may lead to decreased pump performance or pump stop and trigger an Impella Failure or Impella Stopped Controller Failure alarm.

Distribution

US distribution to AR, AZ, CA, CO, CT, FL, GA, ID, MA, MI, MO, MS, NJ, NY, OH, PA, TX. International distribution to Canada, Germany, Great Britian, Italy, Netherlands, Norway.

Lot / Code Info

1. Product Code: 0042-0000-US; GTIN: 813502010022; Serial Numbers: IC1026, IC1029, IC1043, IC1046, IC1040, IC1047, IC1088, IC1089, IC1093, IC1110, IC1167, IC1221, IC1225, IC1222, IC1258, IC1250, IC1272, IC1293, IC1320, IC1317, IC1353, IC1354, IC1355, IC1375, IC1377, IC1403, IC1446, IC1450, IC1452, IC1461, IC1496, IC1501, IC1508, IC1515, IC1516, IC1517, IC1518, IC1530, IC1534, IC1535, IC1539, IC1538, IC1531, IC1533, IC1540, IC1532, IC1551, IC1552, IC1550, IC1553, IC1646, IC1756, IC2255, IC2422, IC2489, IC3013. 2. Product Code: 0042-0000-US-L; GTIN: 813502010022; Serial Number: IC1281. 3. Product Code: 0042-0000-CA; GTIN: 813502011272; Serial Numbers: IC1543, IC1544. 4. Product Code: 0042-0000-EU; GTIN: 813502011289; Serial Numbers: IC1350, IC1383, IC1546, IC1548, IC1547, IC1549, IC1775, IC2243, IC4002. 5. Product Code: 0042-0000-UK; GTIN: 813502011296; Serial Numbers: IC1519, IC1542. 6. Product Code: 0042-0000-UK-L; GTIN: 813502011296; Serial Number: IC1493.

Root Cause

Device Design

Action Taken

Abiomed notified consignees via letter on about 08/20/2025. Consignees were instructed to quarantine and cease use of all affected units, contact Abiomed Field Service team to initiate the remediation process, complete and return the business reply form with the assistance of your Abiomed Representative, forward the recall notification to any applicable personnel and notify customers if affected units were further distributed or transferred. Abiomed also posted a version of the recall notification letter on their website.