NOA Medical Industries Behavioral Health Hospital bed side rails, Product number 1055050GRY-BOX

This recall is currently active, issued January 10, 2025. It was issued by Noa Medical Industries Inc.

Company
Noa Medical Industries Inc
Recall Initiated
November 5, 2024
Posted
January 10, 2025
Recall Number
Z-0838-2025
Quantity
34 units
Firm Location
Washington, MO
Official Source
View on FDA website ↗

Reason for Recall

Red button used to engage side rails can break or become stuck making it difficult to engage the side rail.

Distribution

US: GA, MI. MA, CA

Lot / Code Info

UDI/DI 00817492012745, Serial Numbers: 2038043534, 2038042756, 2038042755, 2038042750, 2038042751, 2038042752, 2038042753, 2038042754, 2038042757, 2038042244, 2038041722, 2038038738, 2038038750, 2038038749, 2038038748, 2038038747, 2038038746, 2038038745, 2038038744, 2038038758, 2038038757, 2038038756, 2038038754, 2038038753, 2038038752, 2038038755, 2038038743, 2038038742, 2038038737, 2038038751, 2038038739, 2038038740, 2038038741, 2038038766.

Root Cause

Device Design

Action Taken

NOA Medical Industries issued an URGENT: VOLUNTARY MEDICAL DEVICE CORRECTION notice to its consignees on 11/05/2024 via email. The notice explained the problem with the device, potential risk, and temporary actions to employ to minimize the hazard until a replacement frame is installed. A replacement side rail will be provided. For questions or concerns, contact: support@noamedical.com (email), https://www.noamedical.com/ (website) or 1-636-239-7600 (telephone), Monday-Friday 7:00AM to 3:30PM, CT.