RecallDepth

PhenoMATRIX

This recall is currently active, issued May 21, 2025. It was issued by Copan Wasp Via Achille Grandi.

Company
Copan Wasp Via Achille Grandi
Recall Initiated
May 6, 2025
Posted
May 21, 2025
Recall Number
Z-1812-2025
Quantity
14
Official Source
View on FDA website ↗

Reason for Recall

AI-powered bacterial culture plate interpretation and workup software does not have Premarket Approval or pre-market notification (510(k)/de novo), so there is potential risk that the colony count and/or morphology information from images may not accurately reflect actual plate and may lead to inaccurate counts and/or morphology characteristics impacting patient diagnosis and subsequent care.

Distribution

US: Nationwide distribution in the states of UT, AR, CA, PA, MD, NC, IL, MA, NJ, KS, TX, NY, NV.

Lot / Code Info

All software versions

Root Cause

No Marketing Application

Action Taken

On 5/6/2025, recall notices were mailed to customers informing them of the following: Ensure the plate images are reviewed against the actual agar plate to verify the assessment you performed by the digital image inspection. Complete and return the response form via email to regulatory.wasp@copangroup.com