PhenoMATRIX
This recall is currently active, issued May 21, 2025. It was issued by Copan Wasp Via Achille Grandi.
- Company
- Copan Wasp Via Achille Grandi
- Recall Initiated
- May 6, 2025
- Posted
- May 21, 2025
- Recall Number
- Z-1812-2025
- Quantity
- 14
- Official Source
- View on FDA website ↗
Reason for Recall
AI-powered bacterial culture plate interpretation and workup software does not have Premarket Approval or pre-market notification (510(k)/de novo), so there is potential risk that the colony count and/or morphology information from images may not accurately reflect actual plate and may lead to inaccurate counts and/or morphology characteristics impacting patient diagnosis and subsequent care.
Distribution
US: Nationwide distribution in the states of UT, AR, CA, PA, MD, NC, IL, MA, NJ, KS, TX, NY, NV.
Lot / Code Info
All software versions
Root Cause
No Marketing Application
Action Taken
On 5/6/2025, recall notices were mailed to customers informing them of the following: Ensure the plate images are reviewed against the actual agar plate to verify the assessment you performed by the digital image inspection. Complete and return the response form via email to regulatory.wasp@copangroup.com