BioFire Respiratory Panel 2.1 plus Kit, REF: 423740, For Filmarray systems

Company
Biofire Diagnostics, Llc
Recall Initiated
April 30, 2025
Posted
May 29, 2025
Recall Number
Z-1877-2025
Quantity
30 kits
Firm Location
Salt Lake City, UT

Reason for Recall

Due to manufacturing error, respiratory panel may result in elevated control failures and/or false negative results.

Distribution

International distribution in the country of Singapore.

Lot / Code Info

Lot# 1475424/ UDI: None

Root Cause

Under Investigation by firm

Action Taken

On April 30, 2025, Biomerieux issued a "Urgent Field Safety Notice" to affected consignees via email. Biomerieux asked consignees to take the following actions: 1. Immediately examine your inventory for the lot identified in this field safety notice. 2. Discontinue use and discard any remaining product from this lot in your possession. 3. As previously confirmed discontinued use of product, please confirm the amount of kits destroyed on the attached Acknowledgement of Receipt Form. 4. If you have further distributed this product, please identify any recipients and notify them at once. 5. Please complete the accompanying Acknowledgement of Receipt Form and return to bioMrieux so that bioMrieux may acknowledge your receipt of this notification.