US AquaA 900H-Typ USA-3 x 208v-Intended for the purification of water to be used for hemodialysis Vivonic Part Numbers: G02020110 Fresenius Part Number: 24-090A-1

Company
Fresenius Medical Care Holdings, Inc.
Recall Initiated
May 16, 2024
Posted
June 13, 2024
Recall Number
Z-2141-2024
Quantity
4 units
Firm Location
Waltham, MA

Reason for Recall

Concentrate Flow was exceeding the maximum 19.9 bar set point. The standard default set point range is between 4 and 18.Error message (F020104, "concentrate pressure alarm limit exceeded") concentrate pressure is too high and stops the dialysis water supply.

Distribution

US Nationwide distribution in the states of AL, IN, AK, WI.

Lot / Code Info

UDI-DI: 04251285200357 Serial Numbers: 2AAV0003 2AAV0029 2AAV0037 3AAV0087

Root Cause

Process control

Action Taken

Fresenius Medical issued Urgent Medical Device Correction letter on 5-16-2024. Letter states reason for recall, health risk and action to take: A Fresenius Medical Care Water Systems Service Specialist will be contacting you to schedule an evaluation of your Water Treatment system and provide feedback on your system. For questions or concerns, please contact Fresenius Medical Care Technical Services at 800-227-2572, OPTION 4. " Please provide this letter to your Biomedical Technician " Report any complaints or adverse events to product.complaints@fmc-na.com or FDA MedWatchathttps://www.fda.gov/safety/medical-product-safety-information/medwatch-forms-fda-safety-reporting. " Provide this Customer Notice to all those who need to be aware within your organization. " Complete and return the attached reply form.

More recalls by Fresenius Medical Care Holdings, Inc.

View all recalls by this company →