Vyaire Medical GmbH, Vyaire Twin Tube sample line 240 cm, Model Number V-707327; The TwinTube sample line 240cm is used in the Vyntus CPX. The TwinTube sample line 240cm is connected to the adapter for DVT. Twin Tube Probenschlauch (707004), which is sold outside the US only. Used in breath analysis for metabolic response study.

This recall is currently active, issued May 8, 2024. It was issued by Vyaire Medical, Inc. Leibnizstr. 7 Hochberg Germany.

Company
Vyaire Medical, Inc. Leibnizstr. 7 Hochberg Germany
Recall Initiated
April 8, 2024
Posted
May 8, 2024
Recall Number
Z-1730-2024
Quantity
649 devices
Official Source
View on FDA website ↗

Reason for Recall

Potential of the nozzle separating during patient use.

Distribution

US nationwide distribution.

Lot / Code Info

UDI/DI 14250892903603, All lots prior to June 2023 or with a serial number before 04I00122 are potentially affected.

Root Cause

Process control

Action Taken

Vyaire Medical issued an URGENT Field Safety Notice to its consignees on 04/08/2024 via email. The notice explained the issue, requested a pull out test for all affected devices to verify the adhesive is adequate prior to use. If the product was further distributed, the party distributing device was directed to notify those to whom the device was sent. For any additional questions or concerns, please contact Vyaire at GMBEMEA- FSCA-RDX-INTL@Vyaire.com.