Alinity m Integrated Reaction Units (IRU)

Company
Abbott Molecular, Inc.
Recall Initiated
February 11, 2022
Terminated
April 14, 2023
Recall Number
Z-0710-2022
Quantity
3063 kits
Firm Location
Des Plaines, IL

Reason for Recall

There is potential for the Reaction Vessels (RVs) detaching from the RV cap during transport by the pipettor robot.

Distribution

US Nationwide distribution in the states of AZ, CA, CO, CT, DC, FL, GA, IL, KY, LA, MA, MI, MO, NC, NJ, NY, OK, RI, SD, TX, WA, and WI.

Lot / Code Info

List Number 09N26-010, Lot Numbers/UDI: 1) 766233 / (01)20884999047864(10)766233(11)210614(240)09N26-010 2) 770433 / (01)20884999047864(10)770433(11)210621(240)09N26-010 3) 771200 / (01)20884999047864(10)771200(11)210712(240)09N26-010 4) 771201 / (01)20884999047864(10)771201(11)210719(240)09N26-010 5) 773642 / (01)20884999047864(10)773642(11)210723(240)09N26-010

Root Cause

Under Investigation by firm

Action Taken

On February 11, 2022, Abbott Molecular issued a Product Recall letter to all customers who received Integrated Reaction Unit (IRU) List Number 09N26-010 Lot Numbers 766233, 770433, 771200, 771201, and 773642 used with the Alinity m System. Consignees must complete and return the included customer reply form. Customers are instructed to follow the Necessary Actions outlined in the Customer Letter: Discontinue use of and discard all remaining inventory of IRU Lots 766233, 770433, 771200, 771201, and 773642. To coordinate replacement, contact your local Molecular Diagnostics at Abbott representative. If RVs are found inside the instrument with no cap and there are associated 1998 errors, please contact an Abbott representative to inspect the instrument. Results generated between the onset of the associated 1998 errors and the identification of the RV with no cap inside the instrument should be considered presumptive. Retesting of samples that yielded positive results during this timeframe should be considered.

More recalls by Abbott Molecular, Inc.

View all recalls by this company →