RecallDepth

VOCSN Patient Breathing Package (Pediatric, Active, Oxygen, Blue) REF PRT-01013-000 with pouch REF PRT-00802-001). Used with the VOCSN Unified Respiratory System which provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.

Company
Ventec Life Systems, Inc.
Recall Initiated
January 9, 2024
Posted
February 23, 2024
Recall Number
Z-1070-2024
Quantity
250 breathing circuits
Firm Location
Bothell, WA

Reason for Recall

The bonded spiral wrap may detach before or during ventilation due to a manufacturing issue. Detachment of the spiral wrap can compromise structural integrity, functionality and/ or performance resulting in occlusions or leaks of the breathing circuit which may cause serious injury or death. The risk of injury or death is increased for ventilator dependent patients.

Distribution

U.S.: CA, CO, FL, HI, IN, MO, NC, and WA O.U.S.: Japan

Lot / Code Info

Box Label - Part Numbers: PRT-01013-000 Pouch Label - Part Number: PRT-00802-001 UDI-DI Code: 00855573007594 Lot Number: 220922 Pack Slip Lot Number: 257337

Root Cause

Process control

Action Taken

On 12/22/2023, the firm sent an URGENT: MEDICAL DEVICE RECALL NOTIFICATION via Direct overnight mail, or FedEx, or email, or hand delivered, to customers informing them that Ventec Life Systems is recalling a specific lot (listed below) of the VOCSN Patient Breathing Package (Pediatric, Active, Oxygen, Blue) since the bonded spiral wrap may detach before or during ventilation due to a manufacturing issue. Customer are instructed to: 1. Operate the ventilator as defined in the Instructions for Use and promptly respond to all alarms activated by the ventilator. 2. Stop use of the affected Patient Breathing Package and replace it with a non-affected product. 3. Inspect current stock and quarantine/segregate any unused affected products to prevent their use. For questions/assistance - contact Ventec Life Systems directly at 844-MYVOCSN (844-698-6276), Monday through Friday between 8:00 a.m. - 5:00 p.m. PDT.

More recalls by Ventec Life Systems, Inc.

View all recalls by this company →