MODULAR CATHCART FRACTURE HEAD HIP BALL, sizes 41mm to 60mm inclusive; metallic femoral head prosthesis, Product No. 136341000, 136342000, 136343000, 136344000, 136345000, 136346000, 136347000, 136348000, 136349000, 136350000, 136351000, 136352000, 136353000, 136354000, 136356000, 136358000, 136360000
- Company
- Depuy Orthopaedics, Inc.
- Recall Initiated
- May 18, 2022
- Recall Number
- Z-1375-2022
- Quantity
- 3488
- Firm Location
- Warsaw, IN
Reason for Recall
The Cathcart Ball has an inherent +5mm offset which is not currently specified in its Instructions for Use or Surgical Technique Guide. The STG has been updated to reflect the clarification that the +5mm offset should be taken into consideration when using a Cathcart head in a construct for hemi-hip arthroplasty procedures.
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Canada, Denmark, Finland, Germany, Greece, Guam, Hong Kong, Ireland, Netherlands, New Zealand, Norway, South Africa, Spain, Sweden, Switzerland, United Kingdom.
Lot / Code Info
All lots until Instructions for Use (IFU) is revised
Root Cause
Labeling design
Action Taken
On June 13, 2022, the firm notified distributed recall letters to customers titled URGENT FIELD SAFETY NOTICE. Customers were instructed to contact their local sales consultant to receive a copy of the revised STG (reference document number DSUS/JRC/0317/2044 Rev B). Product is not being removed from the field.