Omnicell IVX Station-For use in the automated preparation of pharmaceutical admixtures and syringes Model Number: 298921373

Company
Omnicell, Inc.
Recall Initiated
March 27, 2023
Posted
May 22, 2023
Recall Number
Z-1617-2023
Quantity
6 units
Firm Location
Warrendale, PA

Reason for Recall

IVX Station does not account for inherent scale inaccuracy or variability, could result in a preparation being passed that is outside of the dose accuracy stated in the product labeling

Distribution

AL, NY

Lot / Code Info

UDI-DI: 00860006285005 Serial Numbers: IVXS000014, IVXS000015, IVXS000016, IVXS000017, IVXS000018. IVXS000019

Root Cause

Under Investigation by firm

Action Taken

Omincell notified Customers via phone on 3/27/2023 and by Urgent Field Action: Mandatory Correction Of Product Labeling Ivx Station (Limited Release) Letter on 3/31/2023 via email. Letter states reason for recall, health risk and action to take: We will analyze specific devices at your Limited Release sites. Upon completion of our analysis updated labeling will be provided and will include the revised dose accuracy range.