DxU 850m Iris Urine Microscopy Analyzer Catalog Number C49513, in-vitro diagnostic device, urine analysis

Company
Beckman Coulter, Inc.
Recall Initiated
April 5, 2023
Posted
May 17, 2023
Recall Number
Z-1597-2023
Quantity
107 units
Firm Location
Miami, FL

Reason for Recall

Beckman Coulter has become aware of an intermittent issue in which the optional flag "Previous Sample Had Sperm" was enabled but not displayed so that a carryover event could have been investigated.

Distribution

Worldwide Distribution

Lot / Code Info

UDI/DI 15099590748647, All analyzers with Windows 10, Software Versions 8.5, 8.5.1 and 8.6

Root Cause

Under Investigation by firm

Action Taken

Beckman Coulter issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 04/05/2023 by mail. The notice explained the issue and the hazard and requested the following actions be taken: "Examine your specimen settings to determine if the optional 'Sperm Present' and 'Previous Sample Had Sperm' flags are enabled. If disabled, no further action is needed by your laboratory. If enabled, follow the actions below If the presence of sperm is identified: Review the previous specimen for the presence of sperm. Follow recommendations under Previous Sample Had Sperm in the iQ200 Series Instructions For Use (IFU) 300-4320CE and 300-4321EE and DxU 850m and DxU 840m Iris IFU (C49320AB) in Chapter 6, Data Review, Flags, Previous Sample Had Sperm, Recommendations. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product(s) listed above to another laboratory, please provide them a copy of this letter."

More recalls by Beckman Coulter, Inc.

View all recalls by this company →