Medtronic SenSight Extension Tunneler Kit, REF B31030; The Extension Tunneler Kit is used to assist in the tunneling and passage of extensions for deep brain stimulation

Company
Medtronic Neuromodulation
Recall Initiated
March 17, 2023
Posted
April 19, 2023
Recall Number
Z-1408-2023
Quantity
1414 units
Firm Location
Minneapolis, MN

Reason for Recall

Medtronic has received reports of SenSight Extension Tunneler kits that contain dual carriers with have been machined on only one side. As a result, the dual carrier cannot be used to pass two extensions simultaneously.

Distribution

Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canary Islands, Czech Republic, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Luxembourg, Netherlands, New Zealand, Portugal, Slovakia, Spain, Sweden, Switzerland, United Kingdom.

Lot / Code Info

UDI/DI00763000426361, Serial Numbers: HG56BR6, HG56C4R, HG56CJAH01, HG56CLE, HG586N7, HG586UM, HG58704, HG58VM0, HG58VQH, HG58VQHH01, HG58VY5, HG58W7H, HG58WWC, HG58WYB, HG58WYBH01, HG58X0E, HG597BH, HG597DH, HG597M8, HG59ABH, HG59AFC, HG59AV6, HG59RY0, HG5A9GT, HG5A9VY, HG5A9VZ, HG5AQH0, HG5AQQE, HG5BQ5Y, HG5BQLR, HG5BQPJ, HG5BTW7, HG5BTZX, HG5BU5W, HG5BU77, HG5EPSB, HG5EQ0C, HG5EQ2A, HG5EQ69, HG5F31N, HG5F338, HG5F3AZ, HG5F3EH, HG5F3G6, HG5F3K5, HG5F3NF, HG5FP0Y, HG5FP3B, HG5FP70, HG5FP70H01, HG5FPBE, HG5HQBM, HG5HQF8, HG5LJTL, HG5LJZQ, HG5LK2J, HG5LK4H, HG5LK5Y, HG5LK7C, HG5LK8P, HG5LK92, HG5LKAL, HG5LKC9, HG5LKEL, HG5LKGT, HG5LR02, HG5LR33, HG5LTWA, HG5LTWP, HG5LTXU, HG5LTYY, HG5LU12; UDI/DI 00763000486518, Serial Numbers: HG59ABH, HG59RY0, HG59SJ4, HG59SNC, HG5A9H3, HG5A9NB, HG5A9VY, HG5BQ5Y, HG5BQAW, HG5BQLR.

Root Cause

Nonconforming Material/Component

Action Taken

Medtronic issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 03/17/2023 via UPS 2-day delivery. The notice explained the issue and the hazard and requested the following actions be taken: Quarantine all unused product from the affected lots and return them as described on the Customer Confirmation Form. Pass the notice to all those who need to be aware within your organization or to any organization where the potentially affected product has been transferred or distributed.