SmartPerfusion, release R1.0, R1.1, R1.1.1, R1.1.5, R1.1.5.1 and R1.1.6

Company
Philips Medical Systems Nederland B.V. Veenpluis 4-6 Best Netherlands
Recall Initiated
February 20, 2023
Posted
April 20, 2023
Recall Number
Z-1416-2023
Quantity
1461 units (236 US, 1225 OUS)

Reason for Recall

There are technical issues related to signal generation and processing, which can lead to inaccurate presentations.

Distribution

Worldwide - US Nationwide distribution.

Lot / Code Info

UDI: (01)00884838088566 SmartPerfusion with Allura Xper or Azurion model number: 722003, 722006, 722008, 722010, 722012, 722013, 722020, 722023, 722026, 722027 722028, 722029, 722035, 722038, 722039, 722058, 722059, 722064, 722067 722068, 722078, 722079, 722223, 722224, 722225, 722226, 722227 and 722228. Software license key: 459801530271 sw key SmartPerfusion

Root Cause

Software design

Action Taken

A recall notification letter was issued February 20, 2023. The Urgent Medical Device Correction letter includes the following instructions for the Customer: " Stop using SmartPerfusion and 2D Perfusion. NOTE: This message does not impact (other parts of) the Philips Allura or Azurion systems or other tools within the Interventional Workspot. The Philips Allura and Azurion systems as well as other Interventional Workspot tools may continue to be used. " Place the Urgent Medical Device Correction Letter with the documentation of the system. " Circulate the notice to all users so that they are aware of the product issue. " Return the reply form attached to the Urgent Medical Device Correction letter to Philips to confirm that the users of the system have reviewed and understood it.

More recalls by Philips Medical Systems Nederland B.V. Veenpluis 4-6 Best Netherlands

View all recalls by this company →