Malosa Core SMILE Pack 2-Beaver-Visitec- A Medical Refractive Kit intended to be used to aid in cataract, corneal, or refractive surgery. Part Number: MMK967

Company
Beaver Visitec International, Inc.
Recall Initiated
November 17, 2021
Recall Number
Z-0410-2022
Quantity
1170 units
Firm Location
City Point Waltham, MA

Reason for Recall

Kit packaging may have small holes compromising product sterility

Distribution

CA, CO, LA, MA, MD, ND, PA, TN, UT

Lot / Code Info

Lot Numbers: 3347922, 3350452, 3354937, 3358047, 3364371

Root Cause

Packaging change control

Action Taken

Beaver Visitec issued Urgent Medical Device Recall notification dated 11/17/21 to the first level consignee. Letter states reason for recall, health risk and action to take: PLEASE perform the following actions: 1. Immediately examine your inventory and quarantine product from these lots that are subject to recall. If you have further distributed this product, please notify your customers by initiating a sub-recall. Consider all potential users of this product in your user supply chain. You are encouraged to use a copy of this recall notification letter when contacting your customers. 2. Complete the enclosed Business Reply Form (BRF) and indicate whether you have affected product to return. 3. Please enter the following information on the enclosed BRF: part number(s), lot number(s) and quantity(ies) of device(s) that are being returned. 4. Return the completed BRF to BVI by one of these methods: " Fax to 866-912-8416 " Email the attached to beavervisitec3570@sedgwick.com 5. Return your product to STERICYCLE using the enclosed UPS pre-paid return label: Stericycle 2670 Executive Drive, Suite A Indianapolis, IN 46241 Attn: Event 3570 If you have questions regarding the return procedure, please contact Stericycle Inc. at 866-625-5325 (Mon.-Fri. 8:00 am - 5:00 pm Eastern Time).

More recalls by Beaver Visitec International, Inc.

View all recalls by this company →