RecallDepth

Operating Table System

Company
Getinge Group Logistics America, Llc
Recall Initiated
February 11, 2021
Recall Number
Z-1296-2021
Quantity
4288 units
Firm Location
Wayne, NJ

Reason for Recall

There is a potential for hydraulic leakage at the inclination function to occur resulting in an unintended Trendelenburg or anti-Trendelenburg movement of the table top depending on patient positioning which may lead to the patient having to be transferred to another OR table causing a delay in the procedure.

Distribution

Domestic: AK, AL, AR, AZ, CA, CO, DE, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV. Foreign: Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Ghana, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Kuwait, Latvia, Lebanon, Liechtenstein, Lithuania, Luxembourg, Lybia, Macedonia, Malaysia, Mauritius, Mexico, Morocco, Myanmar, Netherlands, New Caledonia, New Zealand, Norway, Oman, Panama, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UAE, United Kingdom, Uzbekistan, Venezuela, Vietnam and Yemen.

Lot / Code Info

Part Number: 118001A0, 118001B0, 118001B1, 118001B2, 118001B3, 118001B4, 118001C0, 118001D0. UDI code: 04046768031500, 04046768031531, 04046768040700, 04046768088351, 04046768088061, 04046768124752, 04046768031555, 04046768031579. Serial Number: 1 3402, 1 1028, 1 1264, 1 401, 3 74, 1 81, 2 1159, 2 - 762.

Root Cause

Device Design

Action Taken

In February, 2021, Maquet GmbH/Getinge, will issue an Urgent Medical Device Correction notice to customers via FedEx 2 Day Delivery for the Magnus 1180 Operating Table due to a potential for hydraulic leakage to occur at the inclination function which may result in an unintended Trendelenburg or anti-Trendelenburg movement of the table top depending on patient positioning. If this were to occur, it may lead to the patient having to be transferred to another OR table causing a delay in the procedure.