RaySert PLUS RSP01. LOT K030. Expiry of 2024-01 - Product Usage: The single use injectors (Model STWOl and Model RSPOl) are intended to be used to compress and insert into the capsular bag only those intraocular lenses that allow the use of these injectors in their approved labelling.

Company
Rayner Intraocular Lenses Limited The Ridley Innovation Centre 10 Dominion Way Worthing United Kingdom
Recall Initiated
March 17, 2020
Recall Number
Z-1753-2020
Quantity
2425 Units.

Reason for Recall

RaySert PLUS RSP01 LOT K030 labeled with an expiry of 2024-01 contained an inner tray that was labeled with an expiry of 2015. In addition to having an expiry of 2015 the LOT number on the inner tray is identified as V001.

Distribution

US nationwide distribution in the states of OH, FL, GA, MD, SC, NJ, PR, IN, ID and MN.

Lot / Code Info

LOT K030.

Root Cause

Labeling mix-ups

Action Taken

Rayner lntraocular notified customers on 03/17/2020 via "Urgent Medical Device recall" letter. The recall letter identified the affected product and lot number. The customers were asked to immediately quarantine and return the product.