Gore DrySeal Flex Introducer Sheath, Catalog #DSF1433, 33cm 14FR, Sterile. The responsible firm on the label is W. L. Gore & Associates, Inc., Flagstaff, AZ.

Company
W. L. Gore & Associates Inc.
Recall Initiated
November 18, 2019
Terminated
May 13, 2020
Recall Number
Z-1962-2020
Quantity
141 devices
Firm Location
Phoenix, AZ

Reason for Recall

The devices were mislabeled with the incorrect product size resulting in 16Fr devices being labeled as 14Fr devices.

Distribution

All distribution is foreign to the following countries: Austria, Belgium, Czech Republic, Denmark, Finland, France, Germany, Greece, Ireland, Italy, Japan, Norway, Spain, Sweden, Switzerland, and the UK.

Lot / Code Info

Serial numbers 21077298, 21077364, 21077365, 21077366, 21077631, 21077350, 21077351, 21077414, 21060888, 21077339, 21077340, 21077341, 21077417, 21077418, 21061626, 21077286, 21077336, 21077337, 21077211, 21077304, 21077305, 21077307, 21077308, 21077309, 21077640, 21077559, 21077563, 21077564, 21077183, 21077346, 21077347, 21077616, 21077617, 21077618, 21077619, 21077419, 21077596, 21077345, 21077620, 21077635, 21077322, 21077602, 21077362, 21077622, 21077623, 21077624, 21077625, 21077626, 21077627, 21077628, 21077348, 21077349, 21077633, 21077644, 21077363, 21077352, 21077634, 21077621, 21077318, 21077319, 21061622, 21077590, 21077591, 21077629, 21077632, 21060886, 21077593, 21077594, 21077274, 21077282, 21077338, 21077636, 21077637, 21077638, 21077639, 21077642, 21077643, 21061625, 21061627, 21077601, 21077608, 21077284, 21077303, 21077320, 21077321, 21077316, 21077415, 21077416, 21077299, 21077300, 21077301, 21077302, 21077369, 21077342, 21077599, 21077600,  21077197, 21077595, 21061623, 21077641, 21077597, 21061624, 21077285,  21077184, 21077185, 21077609, 21077436, 21077437, 21077592, 21077190, 21077191, 21077192, 21077193, 21077194, 21077195, 21077196, 21077206, 21077207, 21077208, 21077561, 21077562, 21077367, 21077368, 21077188, 21077630, 21077186, 21077187, 21077189, 21077310, 21077317, 21077334, 21077353, 21077605, 21077343, 21077344, 21077177, 21077178, 21077179, 21077180, 21077181, and 21077598.

Root Cause

Labeling Change Control

Action Taken

The recalling firm issued Field Safety Notices dated 11/19/2019 with FSN Ref and FSCA Ref numbers of #0000000-11152019-001-R titled "Mislabelled 16Fr Introducer Sheaths Labelled as 14FR (3 lots)" issued via email and overnight mail. Return of the devices was requested.