Neurosign 100, Product Number 9883-01. Intraoperative Nerve Monitor - Product Usage: indicated for use in stimulating and monitoring cranial motor nerves, especially the facial nerve (VII), during ENT surgical procedures.

Company
The Magstim Company Limited Spring Gardens Whitland United Kingdom
Recall Initiated
December 8, 2020
Terminated
February 16, 2023
Recall Number
Z-0869-2021
Quantity
7 units

Reason for Recall

A small number of Neurosign 100 Intraoperative Nerve Monitors may have been dispatched without being configured appropriately for the supply mains within the target Country. The specific deficiency includes: 1. Onboard voltage selector set to 240V (European Standard), versus 120V requirement for the United States and Canada. 2. Fuses installed in the Power Entry Module (PEM) 2xT315mAL (240V Standard), versus the 2xT630mAL requirement for a 120V mains supply.

Distribution

MN, NC, and NY

Lot / Code Info

Serial Numbers: 2507, 2511, 2514, 2515, 2522, 2527, 2533, 2534, 2537

Root Cause

Process control

Action Taken

The firm initiated the recall by email on 12/08/2020. The letter explained the issue, requested the consignee examine the device to confirm that the voltage setting and the fuse ratings are correct.. Should there be a mismatch, the consignee is to discontinue use and notify Magstim Service Department at service@magstim.com or by telephone +44 (0) 199401241111. The firm will send a service pack for the instrument.