RecallDepth

Computed Tomography X-ray System, Model: uCT 550, GTIN: 06971576831036 Intended use The uCT Computed Tomography X-ray System uCT 530/uCT 550 is a computed tomography X-ray system intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck and vascular).

Company
Shanghai United Imaging Healthcare Co., Ltd. No. 2258 Chengbei Rd, Jiading Ind. Shanghai China
Recall Initiated
May 23, 2020
Posted
July 20, 2020
Recall Number
Z-2641-2020
Quantity
3

Reason for Recall

Two issues were identified with the computed tomography x-ray system including a service function which may cause false marking of a bad channel resulting in ring artifacts, and potential intermittent scout scanning interruption due to occasional angle signal drift.. If these problems occur, it may be necessary to rescan the patient resulting in an additional dose of radiation and the possible need for additional contrast medium.

Distribution

US Distribution to state of: TX

Lot / Code Info

Serial Numbers: 353020, 353021, and 353023 with software version R001.11.0.729741-Re-1100

Root Cause

Software design

Action Taken

On 05/23/2020, Customer Notification Letters dated 5/22/2020 were emailed to customers. Customers were asked to do the following: For safety issue 1: If you observe image artifacts, please reboot the system and continue the reconstruction. If the image artifacts persist, please contact the recalling firm's service support team or call the Customer Contact Center at (855) 221-1552 (toll free) or (832) 699-2799 (direct). For safety issue 2: If the scout scanning is interrupted, try repeating the scout scanning. If problem persists, please contact the recalling firm's service support team or call the Customer Contact Center at (855) 221-1552 (toll free) or (832) 699-2799 (direct). Customers were asked to notify all users of the affected products via Customer Notification Letter. On 8/30/2024, a correction notice was sent identifying an additional serial number.

More recalls by Shanghai United Imaging Healthcare Co., Ltd. No. 2258 Chengbei Rd, Jiading Ind. Shanghai China

View all recalls by this company →