RecallDepth

Sterile Disposable Connectors ASY S/1/4 TO 3/16 RED 24 REF/ GTIN for Insert Label:050514000/00803622123269

This recall has been terminated (originally issued April 24, 2020).

Company
Livanova Usa Inc.
Recall Initiated
April 24, 2020
Terminated
February 17, 2022
Recall Number
Z-2450-2020
Quantity
96 units
Firm Location
Arvada, CO
Official Source
View on FDA website ↗

Reason for Recall

The firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a specific production period.

Distribution

US - AK, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WV OUS- Spain (Hospital de Barsuto, Bilbao, Spain)

Lot / Code Info

LOT 2004300006

Root Cause

Packaging process control

Action Taken

The firm sent an URGENT MEDICAL DEVICE REMOVAL letter to affected customers on 04/24/2020. The purpose of this medical device correction is to: 1. Advise the customers of the present issue and not to use the affected devices; 2. Providing information to the user on product disposal or return and replacement; 3. Inform affected customers that LivaNova will coordinate replacement for the unused products.

More recalls by Livanova Usa Inc.

View all recalls by this company →