Maestro Orthodontic Bracket REF711-206-10 QTY. 10 Rx Only (01)00190707016517 (10)112233 (422)484

This recall has been terminated (originally issued April 2, 2019).

Company
Ortho Organizers, Inc.
Recall Initiated
April 2, 2019
Terminated
March 24, 2022
Recall Number
Z-2477-2020
Quantity
104 packages
Firm Location
Carlsbad, CA
Official Source
View on FDA website ↗

Reason for Recall

Orthodontic Bracket is mislabeled

Distribution

13 US Consignees in FL, GA, CA, MI, ID, TX 1 = Italy 1 = Japan 1 = Greece 1 = Saudi Arabia 1 = Russia 1 = Czech Republic

Lot / Code Info

Maestro Orthodontic Bracket REF711-206-10  (01)00190707016517 (10)112233 (422)484 LOT#s: 898430 & 888666

Root Cause

Process control

Action Taken

The firm notified their consignees of the recall by letter sent out 1st class mail on 4/2/2019. The recall was initiated due to a mislabeling error. The Upper Right bracket was erroneously color coded as an Upper Left brackets and then comingled within the same bag. The firm is asking recipients of affected product to please contact Customer Service at 1-800-547-2000 to receive an RMA number. The RMA will allow for a quick return and replacement or credit. The firm is requesting that distributors and physicians complete and E-mail the acknowledgment form back to the firm in order to confirm receipt of the field correction letter. Additionally they are asking authorized distributor,to identify those customers that may have been shipped the affected product and contact those customers to inform them of the issue within fortyeight (48) hours of receipt of this field correction notification.

More recalls by Ortho Organizers, Inc.

View all recalls by this company →