RecallDepth

SoundWave Professional Suite Software 3.2, Ref CI-6055-014, SPN 058-023-32085, UDI: (01)07630016853227, Logical SoundWave Professional Suite 3.2, CE 0123 - Product Usage: Soundwave 3.2 is used by healthcare professionals to program Advanced Bionics sound processors for use with Advanced Bionics cochlear implants, a device is intended to restore a level of auditory sensation to children with profound sensorineural deafness via electrical stimulation of the auditory nerve.

Company
Advanced Bionics, Llc
Recall Initiated
April 4, 2019
Terminated
June 27, 2022
Recall Number
Z-1441-2020
Quantity
364 versions of software
Firm Location
Valencia, CA

Reason for Recall

The manufacturer received complaints that customers were attempting to install the fitting software using the supplied USB drive, an error message was received, and installation failed. The failed installation has not impact to the current version of the software, and users are able to continue programming devices with the current software version.

Distribution

US Nationwide distribution in the sates of AK, AL, AR, AZ, CA, CO, DE, FL, GA, HI, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, OH, OK, PA, SC, SD, TN, TX, VA, VT, WI

Lot / Code Info

Lot #s:  663183 663188 663190 663858 663185 663184 663187 663186 664783 663860 664784 663859 669701 669100 669702

Root Cause

Software in the Use Environment

Action Taken

On 4/4/2019, the recalling firm sent a letter to its customers via FedEx asking that the customer return the corrupted USB drive to the firm.

More recalls by Advanced Bionics, Llc