TriMed Locking Drill Guide, 2.3mm, REF GUIDELCBS-2.3, nonsterile, Rx. The firm name on the label is TriMed Inc., Santa Clarita, CA.

Company
Trimed Inc.
Recall Initiated
September 28, 2018
Recall Number
Z-0356-2020
Quantity
1,042 drill guides
Firm Location
Santa Clarita, CA

Reason for Recall

The epoxy on Apex manufactured epoxy-coated instruments can pop off.

Distribution

Distribution was nationwide. There was military distribution but no government distribution. Foreign distribution was made to Australia, China, South Africa, and Sweden.

Lot / Code Info

Lot numbers 3168, 3464, 3660, 3963, and 4216.

Root Cause

Nonconforming Material/Component

Action Taken

The recalling firm issued letters via email on 9/28/2018 explaining the reason for recall and requesting consignees to identify potentially-affected product and quarantine it until a TriMed representative is able to conduct an inspection.