TriMed Hex Driver 1.2, 55mm AO Cannulated, REF HXDRIVR-1.2C AO, nonsterile, Rx. The firm name on the label is TriMed Inc., Santa Clarita, CA.
- Company
- Trimed Inc.
- Recall Initiated
- September 28, 2018
- Recall Number
- Z-0357-2020
- Quantity
- 467 hex drivers
- Firm Location
- Santa Clarita, CA
Reason for Recall
The epoxy on Apex manufactured epoxy-coated instruments can pop off.
Distribution
Distribution was nationwide. There was military distribution but no government distribution. Foreign distribution was made to Australia, China, South Africa, and Sweden.
Lot / Code Info
Lot numbers 91500604 and 91600017
Root Cause
Nonconforming Material/Component
Action Taken
The recalling firm issued letters via email on 9/28/2018 explaining the reason for recall and requesting consignees to identify potentially-affected product and quarantine it until a TriMed representative is able to conduct an inspection.