AssayTip/AssayCup Tray-used on the cobas e 801 module which is a part of the cobas c 8000 MODULAR Analyzer Series (IVD for in vitro determination of analytes in human body fluids) Catalog Number: 5694302001

This recall has been terminated (originally issued August 13, 2019).

Company
Roche Diagnostics Operations, Inc.
Recall Initiated
August 13, 2019
Terminated
June 30, 2022
Recall Number
Z-0044-2020
Quantity
935 units
Firm Location
Indianapolis, IN
Official Source
View on FDA website ↗

Reason for Recall

AssayTips part of the AssayTip/AssayCup Tray Tips used on the cobas e 801 module- may have an Abnormal Internal Structure and lead to Incorrect patient results

Distribution

Nationwide

Lot / Code Info

Lot Numbers: 18636170 18665170 18667170 18668170 18215274 18245274 18250274 18251274 18252274 18253274 18265274 18271274 19209274

Root Cause

Process change control

Action Taken

Roche Diagnostics issued Urgent Medical Notification letter to Distributor (TP-00744) and end users( TP-00741) on 8/13/19. Letter identifies problem, health risk and Actions required per the UMDC: " Check to see if you have any of the affected lot number(s) of AssayTips/AssayCup Trays on your cobas e 801 module or in your laboratory stock. (See pictures for lot number locations on packaging.). " Stop using the affected lot number(s) of AssayTips/AssayCup Trays immediately. " Discard the affected lot number(s) of AssayTips/AssayCup Trays according to your local waste management guidelines. To receive replacement product, fill out the attached fax form and fax or email it according to the instructions on the form. Contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions