Vigilant Agilia, Vigilant Drug'Lib, REF Z073476, versions 1.0 and 1.1 Product Usage: Vigilant Drug Lib is Dose Error Reduction Software that, when used in combination with Volumat MC Agilia infusion pumps, is intended to reduce drug administration errors at the bedside. Vigilant Drug Lib Agilia is a software application running on a computer and intended to configure and generate multiple device configurations, drug libraries, care areas and data sets that can then be uploaded to the Volumat MC Agilia infusion pumps. Intended patient population: The Vigilant Drug Lib software must only be used to create drug libraries for patients whose weight is 250g to 250 kg (0.6 lbs. to 551 lbs.).

Company
Fenwal Inc
Recall Initiated
June 24, 2019
Posted
August 12, 2019
Terminated
March 6, 2023
Recall Number
Z-2065-2019
Quantity
83 units
Firm Location
Lake Zurich, IL

Reason for Recall

The firm is correcting four software anomalies and KVO (keep vein open) end of infusion alarm priority in Software versions 1.7 and 1.9a of Volumat MC Agilia Software and versions 1.0 and 1.1 of Vigilant DrugLib

Distribution

US Nationwide Distribution

Lot / Code Info

All serial numbers distributed in the US

Root Cause

Other

Action Taken

The firm initiated the recall by letter on 06/24/2019. The letter describes the issues and provided supplemental instructions to be implemented until the new software is installed. The firm issued a press release on 06/28/2019.