Raindrop Near Vision Inlay, REF: RD1-1, Sterile, Rx Only, Manufacturer ReVision Optics, Inc.

This recall has been terminated (originally issued March 5, 2019).

Company
Rvo 2.0, Inc
Recall Initiated
November 13, 2018
Posted
March 5, 2019
Terminated
March 14, 2024
Recall Number
Z-0865-2019
Quantity
2869
Firm Location
Aliso Viejo, CA
Official Source
View on FDA website ↗

Reason for Recall

During a post approval study the firm found there was an increased risk of corneal haze following the implantation of this device.

Distribution

U.S.: NV, GA, NY, MN, TX, ND, AZ, VA, SD, IN, MA, MO, CA, NC, UT, IL, NH, OH, OK, FL, NJ, TN, MI, PA, HI, ID, KY, NE, AL, LA, CO, MD, MS, OR, WI, AR, WA; No OUS (Foreign) Distribution.

Lot / Code Info

All lots of unused product

Root Cause

Unknown/Undetermined by firm

Action Taken

On 11/13/2018, the firm started emailing Urgent Medical Device Recall notices and acknowledgment forms to consignees. This recall notice was intended for 1) patients implanted with the Raindrop Inlay and 2) eye care providers. Eye care providers were asked to cease further implantation and to return unused affected devices to the recalling firm. Eye care providers were informed that patients implanted with affected devices should continue to see their eye care providers on a regular basis, preferably every 6-months and no less often than every 12 months. Also, patients who have had their devices explanted should maintain a follow-up schedule with their provider. The firm commits to physicians to support the patient's best interests and to continue the FDA post approval study to monitor the device's ongoing performance and safety. The firm requests that all consignees complete and return, acknowledgement forms, regardless of whether they do or do not have unused affected product to return. Acknowledgement forms should be returned via email to Jennifer.Luong@opticsmedical.com. Customers with questions are encouraged to call 949-330-6511.