AMSCO 3000 Series Washer/Disinfector Model # 3052

Company
Steris Corporation
Recall Initiated
December 20, 2018
Terminated
July 22, 2019
Recall Number
Z-0995-2019
Quantity
11
Firm Location
Mentor, OH

Reason for Recall

The software in the systems may not process the cycle originally intended. This could result in damage to the various medical devices and instruments processed in the washer/disinfector or potential for inadequate cleaning or disinfection.

Distribution

The products were distributed to the following US states: AL, AR, CA, CO, FL, GA, IL, LA, MA, MD, ME, MI, NC, NJ, NM, NY, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, and WV. The products were distributed to the following foreign countries: Bahrain, Brazil, Canada, China, Saudi Arabia, South Korea, and Taiwan.

Lot / Code Info

Serial Numbers 3602014006 3631618006 3628818009 3629518008 3633018010 3626718010 3630918008 3630918010 3617717013 3620016019 3623516006

Root Cause

Software Manufacturing/Software Deployment

Action Taken

Urgent Device Field Correction notification letters dated 12/20/18 were sent to customers.