Accelerate PhenoTest BC kit, Ref #10101018. The firm name on the label is Accelerate Diagnostics, Tucson, AZ. A multiplexed in-vitro diagnostic test intended for use with the Accelerate Pheno system.

Company
Accelerate Diagnostics Inc
Recall Initiated
March 14, 2018
Terminated
January 17, 2023
Recall Number
Z-1529-2018
Quantity
95 kits
Firm Location
Tucson, AZ

Reason for Recall

The kits may produce an elevated rate of false positive calls for the Staphylococcus aureus identification probe (SAU).

Distribution

Distributed to IL, TX, CA, SC, and DE.

Lot / Code Info

Lot numbers 2880A, 2885A, 2886A, and 2887A, all with exp. date of 9/18/2018.

Root Cause

Under Investigation by firm

Action Taken

The recalling firm contacted all customers on 3/14/2018 via telephone, and a follow-up e-mail was also sent containing a technical bulletin #TB000079, entitled Product Correction Notice, dated March 14, 2018. The notice identified the affected devices and the reason for the recall. Customers were asked to please discontinue use of and dispose of all remaining kits from the affected lots, and document the disposal by completing the 'Certificate of Kit Disposal.' Replacement kits will be provided.